• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. H/W INTERNALBRACE LGMNT AUGMNT REPR KIT; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. H/W INTERNALBRACE LGMNT AUGMNT REPR KIT; FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE Back to Search Results
Model Number H/W INTERNALBRACE LGMNT AUGMNT REPR KIT
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/24/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 09/23/2021, it was reported by a sales representative via phone that when the surgeon was drilling using the ar-8978-cp implant system, he noticed that the kit only contained qty.2 on the 3.5 drill bits.The surgeon used the 3.5 drill bit and when inserted the implant onto the patient, the implant pulled out of the bone as it had no interference.This was discovered during use in a ucl internal brace procedure on (b)(6) 2021.The case was completed by using an ar-1941psv, 3.9 mm knotless corkscrew.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
H/W INTERNALBRACE LGMNT AUGMNT REPR KIT
Common Device Name
FASTENER, FIXATION, NON-DEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key12616680
Report Number1220246-2021-03787
Device Sequence Number3473388
Product Code MBI
UDI-Device Identifier00888867286801
UDI-Public00888867286801
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K150648
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberH/W INTERNALBRACE LGMNT AUGMNT REPR KIT
Device Catalogue NumberAR-8978-CP
Device Lot Number13077046
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2021
Initial Report FDA Received Date10/12/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-