Model Number IPN000320 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
There was no patient involvement.It was reported that the "pump doesn't switch on".There is no information on the event.As a result, the on/off switch has been replaced and the pump was returned to service.
|
|
Manufacturer Narrative
|
Qn#(b)(4).Teleflex received the device for investigation.The reported complaint of "pump doesn't switch on" is confirmed.The terminal between 2a and 2b was noted to be very intermittent which resulted in the reported problem.The root cause of how the terminal became intermittent is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
|
|
Event Description
|
There was no patient involvement.It was reported that the "pump doesn't switch on".There is no information on the event.As a result, the on/off switch has been replaced and the pump was returned to service.
|
|