• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000320
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
There was no patient involvement.It was reported that the "pump doesn't switch on".There is no information on the event.As a result, the on/off switch has been replaced and the pump was returned to service.
 
Manufacturer Narrative
Qn#(b)(4).Teleflex received the device for investigation.The reported complaint of "pump doesn't switch on" is confirmed.The terminal between 2a and 2b was noted to be very intermittent which resulted in the reported problem.The root cause of how the terminal became intermittent is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event Description
There was no patient involvement.It was reported that the "pump doesn't switch on".There is no information on the event.As a result, the on/off switch has been replaced and the pump was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key12617264
MDR Text Key275956484
Report Number3010532612-2021-00304
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902051715
UDI-Public30801902051715
Combination Product (y/n)N
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN000320
Device Catalogue NumberIAP-0500
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2021
Date Manufacturer Received11/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-