• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 4000 ML; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 4000 ML; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX15RE40
Medical Device Problem Code Output below Specifications (3004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the oxygenator presented a low oxygen transfer rate.Per facility, during 15 minutes of bypass, gases with 64 po2 - temperature 35ºc - fio2 60% - oxygen increased to 100% impact on pio2 340 gases.No known impact or consequences to patient.The product was not changed out.The surgery was completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 12, 2021.Upon further investigation of the reported event, the following information is new and/or changed: d9 (device availability - added date returned to manufacturer).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date); g3 (date received by manufacturer); g6 (indication that this is a follow-up report); h2 (follow-up due to additional information and device evaluation); h3 (device evaluated by manufacturer); h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 213, 67).Type of investigation: #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 213 - no device problem found.Investigation conclusions: 67 - no problem detected.The sample was visually inspected and found no break or other anomaly that could lead to the decrease in the gas transfer performance.The actual sample was tested for its gas performance according to the factory inspection protocol.The investigation results revealed that actual sample was the normal product with no issue in the gas performance.In addition, no problem as found in the manufacturing records.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLLOW FIBER OXY WITH 4000 ML
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key12617388
Report Number1124841-2021-00233
Device Sequence Number1321061
Product Code DTZ
UDI-Device Identifier00699753450158
UDI-Public(01)00699753450158
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K153213
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model Number3CX*RX15RE40
Device Catalogue NumberN/A
Device Lot NumberYP14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2021
Supplement Date Received by Manufacturer11/29/2021
01/15/2022
Initial Report FDA Received Date10/12/2021
Supplement Report FDA Received Date12/15/2021
01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age41 YR
Patient SexUnknown
Patient Weight64 KG
-
-