• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANN CONNECTING SCR F/PERCUTAN INSTRUMENTS FOR NAILS-EX; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANN CONNECTING SCR F/PERCUTAN INSTRUMENTS FOR NAILS-EX; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 03.010.404
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, during the insertion of the nail, the targeting arm, static arm screw was not able to insert in the nail so it had to be removed and insert manually.Upon removal of the insertion handle from the nail, the surgeon could not get the insertion bolt out of the nail.Hammer was used to remove the insertion handle.Procedure outcome and patient status are unknown.This report is for one (1) cann connecting scr f/percutan instruments for nails-ex.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: part: 03.010.404, synthes lot: u357728, supplier lot: u357728, release to warehouse date: april 28, 2020, supplier: (b)(4), no nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Visual inspection: the cann connecting scr f/percutan insts nl was received at us customer quality (cq).Visual inspection of the complaint device showed showed only surface scratches due to the usage of the device.Functional test: the device was able to be assembled with the returned insertion handle for suprapatellar.No issue was observed.A full functional assessment was not able to be performed as all applicable mating component(s) were not returned.Dimensional inspection: measured dimension: shaft od = conform, head od = conform.Document/specification review: current and manufactured revisions were reviewed.No design issues or discrepancies were identified.Investigation conclusion: this complaint is not confirmed as there was no defect observed that could have contributed to the reported allegation.No root cause could definitively be determined for the reported complaint condition.No new, unique, or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventive action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANN CONNECTING SCR F/PERCUTAN INSTRUMENTS FOR NAILS-EX
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key12617938
Report Number2939274-2021-06010
Device Sequence Number1415077
Product Code FZX
UDI-Device Identifier10886982068569
UDI-Public(01)10886982068569
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.010.404
Device Catalogue Number03.010.404
Device Lot NumberU357728
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/14/2021
Supplement Date Received by Manufacturer11/09/2021
Initial Report FDA Received Date10/12/2021
Supplement Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AIMING ARM FOR SUPRAPATELLAR; INSERTION HANDLE FOR SUPRAPATELLAR; UNK - NAILS
Patient Age31 YR
Patient SexUnknown
-
-