Model Number 71434487
Medical Device Problem Codes Material Deformation (2976); Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after total knee arthroplasty was performed on an unspecified date, a revision surgery was conducted on (b)(6) 2021 due to implant wear.During revision, it was observed that a legion hinge 4.75mm hex torque wrench adaptor was stripped under normal wear and tear.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports a revision of a competitor prosthesis after a total knee arthroplasty due to implant wear and the adaptor was reportedly ¿stripped under normal wear and tear.¿ the procedure was reportedly completed with a backup smith and nephew device without a surgical delay.The date of the primary surgery was not provided.No additional clinically relevant supporting information has been provided to assist with the clinical investigation.The patient is reportedly ¿doing well.¿ the patient impact beyond the use of a back-up device could not be determined.No further clinical assessment can be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Additional Manufacturer Narrative
After further assessment of the information gathered by the manufacturer, it was identified that this event should be re-evaluated for mdr reporting.The original information stated that, during a revision surgery of a non-s+n device, it was observed that a legion hinge 4.75mm hex torque wrench adaptor was stripped due to normal wear and tear.Surgery was resumed, without any delay, using a smith & nephew back-up device.Patient was not harmed as a consequence of this issue.However, after further clarification and information received, it was determined that the issue does not meet the criteria to be reportable, since patient was not injured and if the device¿s malfunction were to recur, it would not be likely to cause or contribute to a death or serious injury.
 
Event or Problem Description
It was reported that, during a revision surgery of a non-s+n device, it was observed that a legion hinge 4.75mm hex torque wrench adaptor was stripped due to normal wear and tear.Surgery was resumed, without any delay, using a smith & nephew back-up device.Patient was not harmed as a consequence of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION HK 4.75 HEX TORQUE WRENCH ADAPT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
MDR Report Key12619440
Report Number1020279-2021-07545
Device Sequence Number11119823
Product Code HRY
UDI-Device Identifier00885556488454
UDI-Public00885556488454
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number71434487
Device Catalogue Number71434487
Device Lot Number12FAB0012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2021
Supplement Date Received by Manufacturer11/19/2021
11/19/2021
Initial Report FDA Received Date10/12/2021
Supplement Report FDA Received Date11/22/2021
11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/18/2012
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-