The pacemaker was returned and analyzed.The memory content demonstrated a normal functionality of the device while implanted and in service.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed in eri mode.There was no indication of a device malfunction.In agreement with the clinical observation, the device interrogation revealed the battery status eri which was activated on 18-jul-2021.The archive data and the programmed parameters were inspected.High right ventricular and left ventricular pacing outputs of 4.4v and 1.0ms were documented in the archive data.This represents a high current program, which results in a faster discharge of the battery.The amount of charge taken from the battery was verified and the battery condition was found to be as expected.Please note, the pacing percentage of the remaining battery capacity is the remaining capacity until eos.In conclusion, the pacemaker was fully functional.The battery status was anticipated.
|