• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER IRELAND INC. IQ 200 URINE ANALYZER; AUTOMATED URINALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER IRELAND INC. IQ 200 URINE ANALYZER; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Model Number IQ200 SELECT 2008 TESTED
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Urinary Tract Infection (2120)
Date of Event 09/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported leukocytes were not detected on a urine sample on the iq200 instrument causing a delay in patient treatment.Based on available information there was a delay on patient treatment for one patient sample.There was a 10 to 12-hour delay on patient treatment.The patient received antibiotic treatment.Erroneous results reported: zero wbc.Corrected result: 197 wbcul, and 122 wbcul.Leukocytes were not detected on a haematuric urine sample on the iq200 instrument causing a delay in patient treatment.No bacteria detected.This analysis was carried out on the urine of a patient admitted to the emergency room because haematuric urine.Detection of a urinary infection by a strip test was performed.The customer stated that the urine had been diluted to 1/26th as recommended because it was very haematic.The sample was then diluted to 1/2.The leukocytes were detected small particles were then detected, so a gram stain was completed, gram-negative bacilli were observed.The patient received antibiotic treatment(phosphomycin) for a simple urinary infection.
 
Additional Manufacturer Narrative
The assignable cause of the event could not be conclusively determined but the resolution was to stop the automatic validation strategy, customer will continue to monitor to ensure there is no recurrence of the issue.Bec internal identifier - (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ 200 URINE ANALYZER
Common Device Name
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
BECKMAN COULTER IRELAND INC.
lismeehan
o¿callaghan¿s mills
co. clare
EI 
MDR Report Key12619983
Report Number2122870-2021-00164
Device Sequence Number4290779
Product Code KQO
UDI-Device Identifier10837461001300
UDI-Public(01)10837461001300(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K022774
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberIQ200 SELECT 2008 TESTED
Device Catalogue Number700-3345
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2021
Initial Report FDA Received Date10/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-