If information is provided in the future, a supplemental report will be issued.A longevity calculation was completed.Device memory was reviewed and no faults or errors were noted.The device did not trigger elective replacement indicator (eri) and the programmer indicated the device was in doo(r) mode.It was also noted that the device ra sense counter was significantly larger than the rv sense count and there was evidence of prolonged high atrial rates.The device was then exposed to simulated heart load conditions, and the pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.Chronic high atrial rates, as occurred with this device, can reduce longevity in devices that are programmed doo(r) with atrial sensing on.Device power consumption increases proportional to the additional processing for sensed atrial events.As a result, the high rate of atrial sensing was determined to be the cause of the longevity being lower than expected.
|