| Model Number |
385101 |
| Medical Device Problem Codes |
Leak/Splash (1354); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported bd q-syte extension set 15 cm was deformed.The following information was provided by the initial reporter: "hub deformation.".
|
| |
|
Event or Problem Description
|
|
It was reported bd q-syte extension set 15 cm was deformed.The following information was provided by the initial reporter: "hub deformation".
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2021-12-02.H6: investigation summary: our quality engineer inspected the photo and unpacked sample submitted for evaluation.The reported issue was confirmed upon inspection of the photo and sample.Bd can confirm the cause of the failure to be related to our packaging process where the component could have been stuck in our multivac machine, or a scrap material was accidentally packaged.A device history record review showed no non-conformances associated with this issue during the production of this batch.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.A quality alert was generated, and all personnel involved in the process were informed of the failure.
|
| |
|
Search Alerts/Recalls
|