Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2021
Event Type  malfunction  
Manufacturer Narrative
Communication/interviews: a getinge field service engineer (fse) has conversed with the customer and was advised that the batteries were not charging. The fse requested for the customer to make sure the console was secured into the units cart to charge the batteries, the customer confirmed that they were not secured in the console, he installed the console to the cart and the charging sequence started. The customer called the fse and cancelled the service call stating that everything was fine and confirmed that it was a user error. A supplemental report will be submitted upon completion of our investigation. The full event site name in is (b)(6). Device not returned to manufacturer.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) battery was not charging. It is unknown the circumstances under which the event occurred. It is also unknown if there was a patient involvement. However, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID, TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12622694
MDR Text Key276121503
Report Number2249723-2021-02332
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/05/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Reuse

-
-