DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: the getinge field service engineer (fse) resolved the issue by replacing the batteries.The fse performed a full pm with calibration, functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full event site name is (b)(6).The initial reporter named in block is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery run test.Specifically, the battery runtime test was 97 minutes which is below specifications.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|