It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs100 unit display was flickering and with demo balloon system showing autofill error.There was no patient involvement, and no adverse event reported.
|
Updated fields: b4, g3, g6, g7, h2, h4, h6(investigation type, investigation findings, component codes & investigation conclusions), h10, h11.Corrected fields: d1, g1(contact person).Fse replaced the display video receiver cable and checked the unit and it was working fine.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.*the contact person at the event site, pushpalaha, is a biomedical engineer.
|