| Model Number |
BNST413 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the implant fractured and was removed and replaced with a new implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual inspection of the returned product identified worn markings due to usage, excessive bone/debris and a fracture on the thread.No pre-existing conditions were noted on the product experience report (per) the reported device was located on tooth #19 and was used for approximately 2 years 2 months.Pictures or x-ray images were not provided.A device history record (dhr)review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.Complainant reported that the implant fractured and was removed and replaced with a new implant.The reported complaint was confirmed.A definitive root cause for this complaint could not be determined.The following sections have been updated: h3: device evaluated by manufacturer: change ¿no' to 'yes.'.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|