Model Number U125
Device Problem Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p), implanted with another manufacturer's left ventricular (lv) lead, triggered a lead safety switch (lss) due to an unspecified out-of-range lv pace impedance measurement.Review of remote monitoring data was unable to determine the specific measurement.Technical services (ts) reviewed programming and troubleshooting options, and it was reported that lv ring to pacemaker impedance measurements were stable in the 450-600 ohm range.This crt-p and lv lead remain in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12625175
MDR Text Key276189107
Report Number2124215-2021-28090
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/04/2022
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number716565
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/13/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
-
-