BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U125 |
Device Problem
Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p), implanted with another manufacturer's left ventricular (lv) lead, triggered a lead safety switch (lss) due to an unspecified out-of-range lv pace impedance measurement.Review of remote monitoring data was unable to determine the specific measurement.Technical services (ts) reviewed programming and troubleshooting options, and it was reported that lv ring to pacemaker impedance measurements were stable in the 450-600 ohm range.This crt-p and lv lead remain in service.No adverse patient effects were reported.
|
|