| Model Number |
557P |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.There was no report of patient harm or injury this issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging black objects in the pipeline related to a cpap device's sound abatement foam.There was no report of patient harm or injury.After the first attempt to have the device and components returned for evaluation, customer could not able to provide the information about device return.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Section h6 is updated in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was previously contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging black objects in the pipeline related to a cpap device's sound abatement foam.There was no report of patient harm or injury.The device was returned to a third-party service center for investigation.Upon receipt, the device was found to have recorded 20375.8 blower hours and 16987 machine or operation hours.During the evaluation of the device at the service center the device log files were successfully downloaded and reviewed in full.No actionable errors were identified in the log files.In addition, there was no evidence of foam particles or degradation.The customer complaint was not confirmed.The device was scrapped.The manufacturer has updated h6 in this report.
|
| |
|
Search Alerts/Recalls
|