• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MINILOC SAFETY INFUSION SET 20G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MINILOC SAFETY INFUSION SET 20G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of asfsf203 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that "(miniloc) leaked after tightening.They are currently using a medlin kit with two positive pressure needles valves, they are in process of putting t." additional information rcvd 09/27/2021: "i do not have specific lot numbers for these incidents, but the leakage has happened with almost every needle (and continues to) since we started using the product approximately one month ago.The cap on the lower port is loose which allows the blood to leak out.We use approximately 30-50 port needles per day.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINILOC SAFETY INFUSION SET 20G X 0.75IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12627894
Report Number3006260740-2021-04353
Device Sequence Number11605411
Product Code FPA
UDI-Device Identifier00801741048043
UDI-Public(01)00801741048043
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K050600
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0682034
Device Lot NumberASFSF203
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/22/2021
Initial Report FDA Received Date10/13/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-