According to the reporter, during stomach stapling on a laparoscopic sleeve gastrectomy, the stapler fired but the staple line from 4 reloads were not aligned and the staples were randomly formed.The staples were not strongly attached to the tissues.New devices were used to continue the case.There was no patient injury.
|
|
D10: tri 2.0 sig60axt 60 xtra thk 15mm (lot#n1f0002y); tri 2.0 sigc60mt 60 med thk cartridge (lot#n0m0543y); unknown egia adapter (sn:unknown).H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the device had damage to the cutting edge of the knife blade consistent with application over an obstruction.Functionally, the cartridge and loading unit were loaded onto a representative manual handle.The cartridge was applied to test media.The test media was transected.The interlock was tested and found to function properly.It was reported that the staples did not form as expected.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|