The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging caused a patient to develop cancer related to a cpap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.Section h6 was updated in this report.
|
|
The dreamstation cpap pro device was returned to the manufacturer's product investigation lab (pil) for further evaluation.The device powered on, and the airflow was confirmed.The service center retrieved and reviewed the device's error logs.There were 2 errors found.The device was inspected externally and internally.A sticky tapelike substance was observed on the top enclosure.A dust like contaminant was also observed on the top enclosure, keypad, control dial, user interface panel, rear panel, bottom enclosure, printed circuit assembly, blower motor, and blower box.A keratin-like substance was observed on the blower box outlet.During the evaluation, evidence of sound abatement foam degradation/breakdown was not observed.Pil used a fitt (foam integrity test tool) and the associated procedure and was unable confirm the presence of degraded sound abatement foam.The philips pil engineer could not confirm the customer's allegation and there was no visible foam degradation.Sections d9, g1, h2, h3, and h6 have been updated in this report.
|