Model Number 1554-02-165
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Fatigue (1849); Foreign Body Reaction (1868); Necrosis (1971); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Dizziness (2194); Distress (2329); Ambulation or Postural Difficulties (2544); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/22/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pinnacle poly on metal medical record received.After review of the medical records the patient was revised to address pain, difficulty with function mobility, metallosis, polyethylene wear, tear, trunnionosis and high level of metal ions in his blood.Operative note reported some fluid, black tissue that was removed.There was significant trunnionosis.Synovium was removed approx.30 minutes.Doi: (b)(6) 2011 , dor: (b)(6) 2020, left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : a worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.H10 additional narrative: added: h6 clinical symptoms code: appropriate term/code not available (e2402) used to capture blood heavy metal increased.
 
Event or Problem Description
Litigation alleges pain, loss of balance, dizziness, difficulty walking, stomach upset, deterioration and failure of the product, elevated metal ions, abrasion of the metal on metal components with the bearing causing metal debris in the body, tissue dehiscence, necrosis, hypertrophic scarring, antalgia, lumbar strain, fatigue, limited adl and emotional distress.Doi: (b)(6) 2011; dor: (b)(6) 2020; unknown hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Appropriate term / code not available ((b)(4)) is utilized to capture blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AML LG STATURE 16.5MM
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12631211
Report Number1818910-2021-22710
Device Sequence Number16999484
Product Code LPH
UDI-Device Identifier10603295058380
UDI-Public10603295058380
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K012364
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1554-02-165
Device Catalogue Number155402165
Device Lot NumberDK5C61000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2021
Supplement Date Received by Manufacturer12/22/2021
01/04/2022
Initial Report FDA Received Date10/14/2021
Supplement Report FDA Received Date12/28/2021
01/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
56X 36 POLY LINER; AML LG STATURE 16.5MM; ARTICULEZE M HEAD 36MM +15.5; PINNACLE 300 ACET CUP 56MM; 56X 36 POLY LINER; AML LG STATURE 16.5MM; ARTICULEZE M HEAD 36MM +15.5; PINNACLE 300 ACET CUP 56MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-