C.R. BARD, INC. (BASD) -3006260740 SITE~RITE 8 ULTRASOUND SYSTEM 20 MM PINPOINT GT PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770602Y
Medical Device Problem Codes Poor Quality Image (1408); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4)showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed - the image quality is so poor that they are having issues starting iv's with it.
 
Event or Problem Description
Per biomed - the image quality is so poor that they are having issues starting iv's with it.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of poor image quality was unconfirmed.The unit is giving a bright and clear image.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM 20 MM PINPOINT GT PROBE
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
shelly gilbert
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key12631479
Report Number3006260740-2021-04380
Device Sequence Number1402675
Product Code IYO
UDI-Device Identifier00801741138201
UDI-Public(01)00801741138201
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K152554
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9770602Y
Device Catalogue Number9770602Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/24/2021
Supplement Date Received by Manufacturer11/30/2021
Initial Report FDA Received Date10/14/2021
Supplement Report FDA Received Date12/03/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-