Model Number N/A
Device Problems Material Deformation (2976); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event Description
It was reported by healthcare professional "when writer de-accessed patient's ivad, it was difficult to remove.Safety feature not able to be used as needle was found to be very bent upon removal."no other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a bent infusion set needle was confirmed.The product returned for evaluation was a one 22ga x 0.75" safestep safety infusion set.The returned product sample was evaluated and the needle was observed to be bent at the exit site from the non-engaged safety and approximately midway along the needle shaft.The following observations were noted during the sample evaluation: usage residue was seen on the sample which proved that the product had experienced at least some use the bend in the needle had occurred away from the needle base which can be caused by device manipulation during/following port access the needle tip was barbed suggesting contact between the needle and port base.A functional test of the safety mechanism found that the safety mechanism would not engage over the needle tip due to the bend within the needle.Force applied at an angle to the needle axis, or if the needle is not inserted perpendicular into the port septum, can lead to bending of the needle shaft.It is advised to verify that the needle length is correct based on port reservoir depth, tissue thickness and the thickness of any dressing beneath the bend of the needle; if too long, needle and/or port may be damaged at insertion.Additionally, avoid excessive manipulation once the needle is in the port.No potential damage related to the manufacturing process was noted on the complaint sample.H3 other text : sample evaluation findings in manufacturer's notes.
 
Event Description
It was reported "when writer de-accessed patient's ivad, it was difficult to remove.Safety feature not able to be used as needle was found to be very bent upon removal.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 20G X 0.75IN
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12631567
MDR Text Key276387520
Report Number3006260740-2021-04382
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741066160
UDI-Public(01)00801741066160
Combination Product (y/n)N
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0029
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Date Manufacturer Received10/29/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-