• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 014; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 014; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U41502H15
Medical Device Problem Codes Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 10/2023) device pending return.
 
Event or Problem Description
It was reported that during the preparation of an angioplasty procedure, the balloon allegedly ruptured.It was further reported that the balloon detached from the catheter while peeling away the protective sheath.The procedure was completed by using another device.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse 014 pta dilatation catheter has returned for evaluation.On the visual evaluation a complete circumferential break at the weld joint was observed, however the balloon was still intact with the inner guide wire lumen of the catheter.The guide wire lumen was noted to stretched and kinks noted.Both the marker bands are present within the inner guide wire lumen.No evidence of balloon rupture noted on the returned device.No specific anomalies noted.No functional testing performed due to the condition of the device returned for evaluation.Therefore, for the investigation of the reported balloon rupture remains inconclusive as no functional testing performed due to the condition of the device.The investigation for the reported detachment was confirmed as a complete circumferential break at the weld joint was observed separating the balloon from the catheter from the proximal joint, however the balloon was still intact with the inner guide wire lumen of the catheter.A definitive root cause for the reported balloon rupture and detachment of balloon could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 10/2023), g3.H11: h6 (method, result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during the preparation of an angioplasty procedure, the pta balloon allegedly ruptured.It was further reported that the balloon detached from the catheter while peeling away the protective sheath.The procedure was completed by using another device.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 014
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12632845
Report Number2020394-2021-01876
Device Sequence Number19695999
Product Code LIT
UDI-Device Identifier00801741053627
UDI-Public(01)00801741053627
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K121856
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU41502H15
Device Catalogue NumberU41502H15
Device Lot NumberCMEY0191
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/20/2021
Supplement Date Received by Manufacturer12/08/2021
Initial Report FDA Received Date10/14/2021
Supplement Report FDA Received Date12/10/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-