COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number LF1844
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the handpiece sealed inadequately, there was re-grasp alarm, and end tone was observed.A similar device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found no notable conditions.Functionally, the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The knife cut of the device was tested on a silicone test strip with acceptable results.The device was activated multiple times on a saline soaked gauze pad with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.No electrical or wiring issues were found.The device sealing performance was tested on porcine kidney tissue.The sealing was adequate when the seal cycle reached the end tone.It was reported that the device produced only a partial seal of the vessel.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition not related t o the reported condition.: the unit had a jaw failure.The heel gap of the jaw was measured as required and was found to be on the low end of the specification.It has been determined that jaws measuring on the low end of specification, were more likely to generate alarm activations (regrasp alarms) during device use.The most likely cause for the additional condition is traced to device design.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key12632849
Report Number1717344-2021-01410
Device Sequence Number4522255
Product Code GEI
UDI-Device Identifier10884521705890
UDI-Public10884521705890
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K162941
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2025
Device Model NumberLF1844
Device Catalogue NumberLF1844
Device Lot Number01870143X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2021
Supplement Date Received by Manufacturer10/28/2021
Initial Report FDA Received Date10/14/2021
Supplement Report FDA Received Date11/18/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-