U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS,INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS,INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number NDX500S15
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Headache (1880)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused a patient to develop headaches.The patient was hospitalized in response to the reported event.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported an allegation of an issue related to sound abatement foam.Additional information was received and section b5 should be reported as: there is no customer information and we cannot reach out to the customer, hence, attempt to have the device and components returned for evaluation and investigation can not be made at this time.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.The patient alleged patient to develop headaches.There was no medical intervention required by the patient.The reported event of patient to develop headaches and its reported severity was reviewed by the manufacture's clinical expert.This event is assessed as not related to the device in this case.Based on the available information, the manufacture concludes no further action is necessary.There was no medical intervention required by the patient.The device has not yet returned to the manufacturer for evaluation.If any additional information is received, a follow up report will be filed.Section(s) b1, b2, has changed related to the complaint changing from the reported adverse event to a product problem.Section h1 has changed to reflect a malfunction.Section h6 health effect- impact code, type of investigation findings and investigation conclusions has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS,INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS,INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key12633497
Report Number2518422-2021-05292
Device Sequence Number4379849
Product Code BZD
UDI-Device Identifier00606959027206
UDI-Public00606959027206
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberNDX500S15
Device Catalogue NumberNDX500S15
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2021
Supplement Date Received by Manufacturer08/01/2023
Initial Report FDA Received Date10/14/2021
Supplement Report FDA Received Date08/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-