Model Number R SERIES |
Device Problem
Mechanics Altered (2984)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device was unable to switch modes.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll united kingdom for evaluation.The customer's report was duplicated and attributed to the main assembly knob.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|