| Model Number |
42335 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a suspected incident where the valve's setting implanted in patient might have been accidentally adjusted by a 3rd party product.Renol care, a company that produces wrist band, commonly used in wards, the wrist band uses a magnetic tool (similar to the valve adjustment magnet tool) to lock/unlock the band.Usually the ward nurses carry the magnet in their pocket and we suspect that the magnet might have a strong magnetic strength, therefore, resulting in the accidental adjustment of the patient valve settings when placed near (within a certain radius) to the patient's head.It was stated the patient was fine, and there was no significant impact to the patient as the issue was detected early.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that the patient setting was noticed to have been adjusted unknowingly.The nurse clinician suspected that it was the magnet from the wrist band as the main cause for the issue as it was a common accessory which ward nursesmcarry along with them.
|
| |
|