Catalog Number 302149 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: unknown.The date received by manufacturer has been used as a default.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported that syringe 10ml ll leaked.The following information was provided by the initial reporter: "syringe rubber stopper will leak".
|
|
Manufacturer Narrative
|
H6: investigation summary two photos were received by our quality team for evaluation.From the photos, leakage was observed past the stopper.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.At the assembly process there is a mechanism control to check the stopper leakage.However, as no sample was returned from further analysis, a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that syringe 10ml ll leaked.The following information was provided by the initial reporter: "syringe rubber stopper will leak".
|
|