Catalog Number 10226
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer reported that a patient had a therapeutic plasma exchange (tpe) procedure on (b)(6) 2021.The procedure used 1000ml fresh frozen plasma+1000ml common plasma.Per the customer, the patient appeared pneumorrhagia, epistaxis and blood clotting function abnormal at (b)(6) 2021.The patient returned to normal after infused with 400ml fresh frozen plasma.On (b)(6) 2021, the operator used saline to replace ac due to the abnormal clotting after previous procedure, 1000ml fresh frozen plasma+1000ml common plasma.After 2 hours of the procedure, the inspection result of blood clotting function was normal.But after 7 hours of the procedure, the inspection result was abnormal again.The exchange set is not available for return because it was discarded by the customer.Patient information and outcome are unknown at this time.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: the run data file (rdf) was analyzed for this event.Review of the rdf associated with this complaint show very high inlet:ac ratios for the procedures.However, according to the complaint the concentration of the ac used differentiates from the standard concentration.Review of the images does not show any clumping or any signs of issues during the procedure.The optia system appears safe to use and operated as intended.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer reported that a patient had a therapeutic plasma exchange (tpe) procedure on (b)(6) 2021.The procedure used 1000ml fresh frozen plasma+1000ml common plasma.Per the customer, the patient appeared pneumorrhagia, epistaxis and blood clotting function abnormal at (b)(6) 2021.The patient returned to normal after infused with 400ml fresh frozen plasma.On (b)(6) 2021, the operator used saline to replace ac due to the abnormal clotting after previous procedure, 1000ml fresh frozen plasma+1000ml common plasma.After 2 hours of the procedure, the inspection result of blood clotting function was normal.But after 7 hours of the procedure, the inspection result was abnormal again.The exchange set is not available for return because it was discarded by the customer.Patient information and outcome are unknown at this time.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Event or Problem Description
The customer reported that a patient had a therapeutic plasma exchange (tpe) procedure on (b)(6) 2021.The procedure used 1000ml fresh frozen plasma+1000ml common plasma.Per the customer, the patient appeared pneumorrhagia, epistaxis and blood clotting function abnormal at (b)(6) 2021.The patient returned to normal after infused with 400ml fresh frozen plasma.On (b)(6) 2021, the operator used saline to replace ac due to the abnormal clotting after previous procedure, 1000ml fresh frozen plasma+1000ml common plasma.After 2 hours of the procedure, the inspection result of blood clotting function was normal.But after 7 hours of the procedure, the inspection result was abnormal again.The exchange set is not available for return because it was discarded by the customer.Patient information and outcome are unknown at this time.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Corrected information is provided in b.3.Investigation: the run data file (rdf) was analyzed for this event.Review of the rdf associated with this complaint show very high inlet:ac ratios for the procedures.However, according to the complaint the concentration of the ac used differentiates from the standard concentration.Review of the images does not show any clumping or any signs of issues during the procedure.The optia system appears safe to use and operated as intended.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer reported that a patient had a therapeutic plasma exchange (tpe) procedure on (b)(6) 2021.The procedure used 1000ml fresh frozen plasma+1000ml common plasma.Per the customer, the patient appeared pneumorrhagia, epistaxis and blood clotting function abnormal at (b)(6) 2021.The patient returned to normal after infused with 400ml fresh frozen plasma.On (b)(6) 2021, the operator used saline to replace ac due to the abnormal clotting after previous procedure, 1000ml fresh frozen plasma+1000ml common plasma.After 2 hours of the procedure, the inspection result of blood clotting function was normal.But after 7 hours of the procedure, the inspection result was abnormal again.The exchange set is not available for return because it was discarded by the customer.Patient identifier was not provided by the customer.The customer confirmed the coagulation results returned to normal ranges.This product is not available within the us, but this report is being filed due to an alleged failure that could occur on a similarly marketed device platform cleared for use by the fda.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in a.2, a.3, a.4, b.5, b.6, e.1, h.6 and h.10.Investigation: the run data file (rdf) was analyzed for this event.Review of the rdf associated with this complaint show very high inlet:ac ratios for the procedures.However, according to the complaint the concentration of the ac used differentiates from the standard concentration.Review of the images does not show any clumping or any signs of issues during the procedure.The optia system appears safe to use and operated as intended.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.According to therapeutic plasma exchange as management of complicated systemic lupus erythematosus (sle) and other autoimmune diseases (aguirre-valencia, et al, hindawi volume 2019, article id 5350960, p.2): the most prevalent diagnosis was sle (n=40, 60.61%).In these patients, the main indication of tpe was diffuse alveolar hemorrhage (dah; n=11, 27.5%) and neuropsychiatric involvement (n=9, 22.5%), including patients with central, peripheral, and psychiatric compromise.The other indications in patients with sle were thrombotic microangiopathy (n=5, 12.5%), catastrophic aps (n=4, 10%), lupus nephritis (n=4, 10%; all had type iv diffuse proliferative nephritis), gastrointestinal involvement (lupus pseudoobstruction; n=1, 2.5%), severe sle (polyserositis and joint swelling; n=1, 2.5%), severe resistance to type b insulin (n=1, 2.5%), evans syndrome (n=1, 2.5%), and kidney allograft rejection (n=1, 2.5%).A disposable lot history search indicated there were no other reported occurrences of adverse reactions on this lot worldwide.Root cause: a root cause assessment was performed for the adverse reactions experienced by both patients.Based on the information provided by the customer, the dlog analysis, and the scientific journal by aguirre-valencia, et al, the adverse reactions experienced by both patients are frequent pathologies associated with systemic lupus erythematosus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Common Device Name
SPECTRA OPTIA EXCHANGE SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key12636572
Report Number1722028-2021-00322
Device Sequence Number4178731
Product Code LKN
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Catalogue Number10226
Device Lot Number2104073130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2021
Supplement Date Received by Manufacturer10/15/2021
01/11/2022
04/07/2022
Initial Report FDA Received Date10/14/2021
Supplement Report FDA Received Date11/08/2021
01/14/2022
04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age15 YR
Patient SexFemale
Patient Weight49 KG
-
-