| Model Number |
M0061921440 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Difficult to Advance (2920); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to a sensor wire and polaris ultra stent during the same procedure.Refer to manufacturer report# 3005099803-2021-05215 for the other associated device information.It was reported to boston scientific corporation that a polaris ultra ureteral stent and a sensor wire was used during ureteroscopy procedure in the kidney performed on (b)(6) 2021.During procedure closure, the technician encountered resistance when attempting to pull the guidewire out and physician instructed her to continue.The guidewire tip tore the entire length of the stent and the coil was detached.Upon removal, it was noted the guidewire corewire tip was broken inside the patient when wire was being pulled out.The procedure was completed with a new polaris ultra stent and sensor wire.There were no patient complications reported as a result of this event.
|
| |
|