• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW ADULT NASAL CANNULA; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW ADULT NASAL CANNULA; CAT Back to Search Results
Model Number OPT844
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The opt844 interface is used to deliver humidified oxygen to patients.The opt844 consists of a lightweight delivery tube which is connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's nares.Method: the complaint opt844 nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photograph provided by the customer, and our knowledge of the product.Result: visual inspection of the provided photograph revealed that the interface was found broken at the right headgear connection point.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.However, the cause of the reported event is most likely due to overtightening of the head strap.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is disposed of.The subject cannula would have met the specification at the time of production.The setup instructions in the user instructions which accompany the opt844 nasal cannula include the following steps: - "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." the user instructions also contain the following warnings/cautions: "do not crush or stretch tube, to prevent loss of therapy." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in (b)(4), that the connector hook of the opt844 optiflow adult nasal cannula was found damaged before patient use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW ADULT NASAL CANNULA
Common Device Name
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key12637273
Report Number9611451-2021-01161
Device Sequence Number5504906
Product Code CAT
UDI-Device Identifier09420012432124
UDI-Public(01)09420012432124(10)2101494757(11)210204
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT844
Device Catalogue NumberOPT844
Device Lot Number2101494757
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2021
Initial Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-