| Model Number |
AMSCO 600 Steam Sterilizer |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer is conducting an investigation as a follow-up to three infections that occurred in three separate patients post procedure.The infections occurred following procedures with the same surgeon.The user facility did not disclose the current patient status or if any medical treatment was administered.We understand the hospital is reviewing all their pre- and post-procedural standard operating procedures.The customer contacted steris asking we perform an evaluation of their equipment in an attempt to rule out the amsco 600 sterilizers as the root cause of the reported infections.The user facility did not disclose the specific instruments that were used on the patients that obtained infections.The user facility reported that biological indicators were present and evidenced passing results.Biological indicators are utilized with every load.A steris service technician arrived onsite to inspect the sterilizers.The technician ran test cycles which completed successfully and confirmed the amsco 600 sterilizers to be operating according to specification.The technician proactively descaled the steam generators and returned the units to service.The technician also inspected the chambers and rack assemblies and found no issues.Our investigation concluded that our amsco 600 sterilizers are operating according to specification.A steris clinical education specialist provided a refresher in-service training to the customer on proper pre-cleaning, processing, and inspection of devices.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility contacted steris requesting functional testing of their three amsco 600 sterilizers.
|
| |
|
Search Alerts/Recalls
|