| Model Number |
ORT 400 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow-up report will be submitted once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a magnetic resonance-guided litt procedure, the area targeted for ablation appeared, prior to ablation, to be 4.5 mm to 8 mm off target.The case was completed after noting the possible image distortion issue and no patient impact or consequence was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The root cause of this event was traced to user error of incorrect localizer settings; the first localizer settings were not zeroed out which led to an error impacting patient registration.The phase shift was reported to be observed prior to ablation therapy, after which coordinate adjustments were made and the procedure continued without further impact.A device malfunction was not identified during the investigation.
|
| |
|