Model Number CLV-S40PRO |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was evaluated at olympus service operation repair center (sorc).As a result of the evaluation, the following was confirmed.Due to the wear of the output socket, there was rattling when connecting the ride guide.The front panel flashed due to sticking due to deterioration of the filter turret and mesh turret.The foot rubber on the bottom was damaged.There was a dent on the switch on the front panel and it was scratched.The amount of light was reduced due to the consumption of the xenon lamp.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus inspected the device at olympus service operation repair center (sorc) and found that the fuse box of the ac power inlet was burnt.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.From the investigation results, the exact cause of the reported event could not be conclusively determined.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
|
|