Model Number CLV-S40PRO
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated at olympus service operation repair center (sorc). As a result of the evaluation, the following was confirmed. Due to the wear of the output socket, there was rattling when connecting the ride guide. The front panel flashed due to sticking due to deterioration of the filter turret and mesh turret. The foot rubber on the bottom was damaged. There was a dent on the switch on the front panel and it was scratched. The amount of light was reduced due to the consumption of the xenon lamp. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus inspected the device at olympus service operation repair center (sorc) and found that the fuse box of the ac power inlet was burnt. There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVISERA PRO XENON LIGHT SOURCE
Type of DeviceXENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12638037
MDR Text Key281068035
Report Number8010047-2021-13193
Device Sequence Number1
Product Code GCT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/25/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCLV-S40PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2021
Was the Report Sent to FDA? No
Date Manufacturer Received11/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2006
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-