| Model Number |
0998-00-3013-92 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Testing of actual/suspected device : a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and found the battery test failed < 137 min of autonomy.The fse resolved the issue by replacing the batteries.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The initial reporter named in is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs100 intra-aortic balloon pump (iabp) failed battery test requiring battery replacement.There was no patient involvement, and no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of production: (3331) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period (b)(6) 2019 through (b)(6) 2021 was reviewed.There were no triggers identified for the review period.
|
| |
|
Search Alerts/Recalls
|