• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 1294-53-246
Medical Device Problem Codes Osseointegration Problem (3003); Migration (4003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 10/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient's femoral sleeve did not ingrow and subsided.Components were removed and new components were placed.This patient has a below knee amputation on the opposite leg so limited weight bearing is not possible likely contributing to lack of fixation.There was also an indication of loosening with the femoral sleeve.Doi: (b)(6) 2021, dor: (b)(6) 2021 right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays and photographs were reviewed, and it is not possible to confirm a depuy's product failure as well as a relation between the adverse event and the reported product.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL FEM SLV FUL POR 46MM
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12638334
Report Number1818910-2021-22779
Device Sequence Number14315744
Product Code JWH
UDI-Device Identifier10603295026365
UDI-Public10603295026365
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1294-53-246
Device Catalogue Number129453246
Device Lot Number714531
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2021
Supplement Date Received by Manufacturer11/29/2021
Initial Report FDA Received Date10/15/2021
Supplement Report FDA Received Date11/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE REV LPS INSRT XXSM 12MM; CONTROL CABLE W/CABLE SLEEVE; LPS DISTAL FEM COMP XXSM RT; LPS FEM TO SLEEVE ADAPTER +5; UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL STEM 75X18MM FLUTED; ATTUNE REV LPS INSRT XXSM 12MM; CONTROL CABLE W/CABLE SLEEVE; LPS DISTAL FEM COMP XXSM RT; LPS FEM TO SLEEVE ADAPTER +5; UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL STEM 75X18MM FLUTED
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexMale
Patient Weight127 KG
-
-