• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US GLOBAL ANCHOR PEG GLENOID 48MM; GLENOID (STEPTECH AND ANCHOR PEG) : SHOULDER GLENOID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US GLOBAL ANCHOR PEG GLENOID 48MM; GLENOID (STEPTECH AND ANCHOR PEG) : SHOULDER GLENOID Back to Search Results
Model Number 1136-42-025
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision of right total shoulder due to poly wear.Original dos was over 20 years ago.Doi: 20 years ago, dor: (b)(6) 2021, right shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative:  if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the adverse consequences were that patient needed revision to reverse total shoulder due to poly wear on the glenoid component.No delay or extended surgery time.This was not a natural progression.The lot number of glenoid was illegible due to poly wear.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination, therefore unavailable for a physical evaluation.However a photograph was provided.Upon visual inspection of the photograph, it was noticed that the photo does not depict the product that is part of the allegation.Therefore, the complaint cannot be confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL ANCHOR PEG GLENOID 48MM
Common Device Name
GLENOID (STEPTECH AND ANCHOR PEG) : SHOULDER GLENOID
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12638518
Report Number1818910-2021-22791
Device Sequence Number3452047
Product Code KWS
UDI-Device Identifier10603295006985
UDI-Public10603295006985
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K981487
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1136-42-025
Device Catalogue Number113642025
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2021
Supplement Date Received by Manufacturer10/14/2021
11/29/2021
Initial Report FDA Received Date10/15/2021
Supplement Report FDA Received Date10/20/2021
11/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLOBAL ADVANTAGE HUM HD 48X21; GLOBAL ADVANTAGE STEM 14 MM; GLOBAL ANCHOR PEG GLENOID 48MM; GLOBAL ADVANTAGE HUM HD 48X21; GLOBAL ADVANTAGE STEM 14 MM; GLOBAL ANCHOR PEG GLENOID 48MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight115 KG
-
-