| Catalog Number |
0010216 |
| Medical Device Problem Codes |
Contamination of Device Ingredient or Reagent (2901); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At this time, no conclusion can be made.Returned for evaluation were four (4) small round white pieces of the alleged foreign material found by the user within the product packaging.Preliminary evaluation finds the returned material is round and plastic in nature.Review of the returned sample is ongoing at this time.When the sample evaluation is completed, a supplemental mdr will be submitted.A review of manufacturing records shows product was manufactured to specification.To date, this is the only complaint reported for this manufacturing lot of (b)(4) units released for distribution in june, 2019.
|
| |
|
Event or Problem Description
|
|
As reported, upon opening the package of the bard/davol ventrio mesh during a procedure on (b)(6) 2021, a "white foreign material" was noted within the inner package.As reported, no damage was noted on the outer package and the mesh was implanted in the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
At this time, no conclusion can be made.Returned for evaluation were four (4) small round white pieces of the alleged foreign material found by the user within the product packaging.Preliminary evaluation finds the returned material is round and plastic in nature.Review of the returned sample is ongoing at this time.When the sample evaluation is completed, a supplemental mdr will be submitted.A review of manufacturing records shows product was manufactured to specification.To date, this is the only complaint reported for this manufacturing lot of (b)(4) units released for distribution in june, 2019.Addendum: this is an addendum to the initial mdr, to document our investigation results.Based on visual and manufacturing process evaluation performed, the origin of the white foreign material could not be matched with a manufacturing generated condition.Based on manufacturing record review, no raw material component or sub assembly generated at the manufacturing site was identified with a similarity that could be related to condition reported.Based on the sample evaluation and the investigation performed, the source of the plastic material could not be determined.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : sample evaluated.
|
| |
|
Event or Problem Description
|
|
As reported, upon opening the package of the bard/davol ventrio mesh during a procedure on (b)(6) 2021, a "white foreign material" was noted within the inner package.As reported, no damage was noted on the outer package and the mesh was implanted in the patient.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|