| Model Number |
IPN913776 |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of the reported failure mode was conducted and 13 devices were taken from current production (425-00 concha neptune lot # 73h210002r) at the manufacturing facility and were inspected.During the test, issue reported "unit not maintaining temp prior to use" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported that the unit is "not maintaining temp".No patient involvement reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the unit is "not maintaining temp".No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn #(b)(4).The sample was returned for evaluation.A visual exam was performed and damage was observed on the rear mounting bracket.No other defects were observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was able to power on and off with no issues.The unit was then prepared for the functional bench test where a water reservoir bottle, an adult breathing circuit (880-36kit), a 382-10 conchasmart column, airflow, and dual temperature probes were connected to the unit.The unit successfully navigated all pre-operational self-tests again and was functioning in real time.The settings were set to full rainout, invasive, at 37 degrees c.The unit functioned without any interruption and was able to heat up to the set temperature of 37 degrees c.The unit was able to maintain the set temperature for approximately 1 hour with no issues.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
|
| |
|
Search Alerts/Recalls
|