The fracture of the guidewire will be reported in mdr 3004742232-2021-00347.Device analysis conclusion: the oad and guide wire were received at csi for analysis.The guidewire was engaged in the oad.Visual examination confirmed that each component was fractured.The distal driveshaft fragment was not received, which also confirms the report that the oad fragment remained in vivo at the conclusion of the procedure.The distal viperwire fragment was received in a separate bag, which does not support the report that the wire fragment was left in vivo.Driveshaft flexing at the weld location can initiate a fatigue failure in an oad, and scanning electron microscopy (sem) analysis identified fatigue striations at the site of the driveshaft fracture.It is hypothesized that this driveshaft underwent excessive flexing near the crown due to spinning in excessive tortuosity or resistance that pushed the driveshaft into a tight bend shape.However, the exact root cause of the fracture could not be determined.Analysis of the device data log revealed stall events.The stall events may be related to the driveshaft fracture, but this could not be conclusively confirmed.Evidence of ductile torsion was observed during sem analysis of the guidewire.Analysis did not determine that the wire fractured due to design or failure to meet specification.Instead, analysis indicated the wire was exposed to extreme and abnormal stress conditions.This is consistent with details reported to csi about the procedure.However, the root cause of the wire fracture could not be conclusively determined.The device history record and material inspection report for the device and wire lot numbers have been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event and the device met material, assembly, and quality control requirements prior to distribution.In the opinion of the physician, the fracture was believed to be due to severe angulation and highly calcified state of the vessel.Csi id: (b)(4).
|
|
Distal to proximal treatment with a diamondback gen 2 coronary orbital atherectomy device (oad) was performed in the mid circumflex (cx) artery.A second treatment was initiated, but the oad stalled.The driveshaft fractured on the proximal side of the crown.The guide wire also fractured.The oad fragment was lodged in the distal left main and proximal cx.A triple wire snare technique was used in an attempt to dislodge the fragment.However, the fragment could not be retrieved and was not visible on imaging at the conclusion of the procedure.The patient was agitated at the conclusion of the procedure, and the reason for the agitation was believed to be from a response to anesthetic.The results of the procedure were otherwise described as "excellent.".
|