MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED HELICAL BLADE 75 - SILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED HELICAL BLADE 75 - SILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.038.375S
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 09/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional procode: ktt.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part # 04.038.375s, synthes lot # h566892, supplier lot # n/a, release to warehouse date: 12 apr 2018, expiration date: 31 mar 2028, manufacturing location: (b)(4), no ncr's were generated during production.Device history review: review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, during an unknown procedure, while inserting a tfna nail, the surgeon perforated the femoral neck with the helical blade.The nail and helical blade were extracted, and reinserted.The helical blade was repositioned in the femoral neck and the surgery was completed successfully.During extraction, one of the prongs of the connection interface was slightly bent; however the nail was successfully reconnected and inserted.There was a fifteen (15) minute surgical delay.The procedure was successfully completed with no patient consequences.This report is for one (1) tfna fenestrated helical blade 75 - sile.This is report 1 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED HELICAL BLADE 75 - SILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12639890
Report Number2939274-2021-06125
Device Sequence Number3413479
Product Code HSB
UDI-Device Identifier10886982099198
UDI-Public10886982099198
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.038.375S
Device Catalogue Number04.038.375S
Device Lot NumberH566892
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/17/2021
Initial Report FDA Received Date10/15/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
10/130 DEG TI CANN TFNA 340/LEFT - SILE; RIA 2 BONE HARVESTING KIT 520MM STERILE
Patient Age48 YR
Patient Weight68
-
-