• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. ANGIOCATH BL 22GA X 1.0IN; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. ANGIOCATH BL 22GA X 1.0IN; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381123
Medical Device Problem Code Free or Unrestricted Flow (2945)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that angiocath bl 22ga x 1.0in had air in the line.The following information was provided by the initial reporter: this is a report about air in line with the use of angiocath.According to the customer's report, when connecting to the a line, air entered.The customer is suspecting that there may be a crack on the catheter or the catheter hub.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 2021-10-25 h6: investigation summary bd received a 22 gauge angiocath device from lot 9329998 for evaluation.A review of the device history record was performed for the reported lot and no quality issues were found during production.Our quality engineer visually inspected the returned unit and observed no visual damage to the unit.Next, the unit was sent for a ct scan to identify if there were any micro-cracks that could not be identified.The ct scan confirmed that there was no damage to the sample that could have contributed to a possible leak or entryway for air.Based off of the visual inspection and ct scan the engineer could not verify the reported claim.Since no defects were found during inspection a definitive root cause could not be determined.H3 other text : see h10.
 
Event or Problem Description
It was reported that angiocath bl 22ga x 1.0in had air in the line.The following information was provided by the initial reporter: this is a report about air in line with the use of angiocath.According to the customer's report, when connecting to the a line, air entered.The customer is suspecting that there may be a crack on the catheter or the catheter hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOCATH BL 22GA X 1.0IN
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer (Section G)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12640273
Report Number9610048-2021-00140
Device Sequence Number4343988
Product Code FOZ
UDI-Device Identifier00382903811236
UDI-Public00382903811236
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2024
Device Catalogue Number381123
Device Lot Number9329998
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2021
Supplement Date Received by Manufacturer12/30/2021
Initial Report FDA Received Date10/15/2021
Supplement Report FDA Received Date01/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-