• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12MM / TI CANN FRN / GT 400MM / RIGHT - STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12MM / TI CANN FRN / GT 400MM / RIGHT - STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 04.033.270S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Complainant part is not expected to be returned for manufacturer review/ investigation.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.The product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot, part: 04.033.270s-us, lot: l984217, manufacturing site: (b)(4), release to warehouse date: 13 august 2018, expiration date: 31 july 2028.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent for a surgery.During the surgery, while attempting to insert recon screws into an unknown femoral recon nail (frn), the surgeon hit the nail with the drill bit.The nail was unstable because the surgeon did not pin both trocars before removing guide wire to drill the first hole.There was no surgical delay.The procedure was successfully completed.There was no patient consequence reported.This complaint involves an unknown number of devices.This report is for (1) 12mm / ti cann frn / gt 400mm / right - sterile.This report is 1 of 10 of (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12MM / TI CANN FRN / GT 400MM / RIGHT - STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12641953
Report Number2939274-2021-06131
Device Sequence Number11616288
Product Code HSB
UDI-Device Identifier10886982271815
UDI-Public(01)10886982271815
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K172157
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.033.270S
Device Catalogue Number04.033.270S
Device Lot NumberL984217
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2021
Initial Report FDA Received Date10/15/2021
Patient Sequence Number1
Patient Age47 YR
Patient Weight87
-
-