|
It was reported that the procedure was to treat a lesion located in the left anterior descending coronary (lad) artery back into the left main (lm).The turntrac wire was jailed behind stent during bifurcation stenting lad back into lm.Both physicians mentioned that the wire was not any more severely jailed than usual and there was no resistance on the wire.When removing the guide wire, it snapped a few centimeters proximal to the radiopaque section.The wire then unraveled, released fragments throughout the patients lad and lm.There was a 2-3 hour period of snaring fragments of the wire out of the lad and lm using the intravascular ultrasound catheter and a non-abbott guide wire.All fragments that were visible on angiogram were removed from the patient.There was no adverse patient sequela.No additional information was provided.
|
|
Device code 2017 - failure to follow steps / instructions.A visual inspection was performed on the returned guide wire and the reported separation and stretched coils was confirmed.The reported entrapment of the device was unable to be replicated or confirmed in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.There was no damage noted to the guide wire during the inspection prior to use which suggests a product quality issue with respect to manufacture, design or labeling did not contribute to the reported difficulties.It should be noted that instructions for use guide wire hi-torque turntrac, ce/fda: carefully observe the instructions under ¿do not¿ and ¿do¿ below.Failure to do so may result in vessel trauma, guide wire damage, guide wire tip separation, or stent damage.If resistance is observed at any time, determine the cause under fluoroscopy and take remedial action as needed.Use the most suitable guide wire for the lesion being treated.Do not push, auger, withdraw, or torque a guide wire that meets resistance.Do not torque a guide wire if the tip becomes entrapped within the vasculature.Do use extreme caution when moving a guide wire through a non-endothelialized stent, or through stent struts, into a bifurcated vessel.Use of this technique involves additional patient risks, including the risk that the wire may become caught on the stent strut.In this case, the technique used during the procedure of not pulling back the wire prior to stent deployment causing the entrapment of the guide wire with the stent resulting in the wire separation, stretched coils and removal of foreign body and delays in the procedure, does appear to have caused or contributed to the reported complaint.The investigation determined the reported entrapment of the device, guide wire separations, stretched coils, unexpected medical intervention, and removal of foreign body resulting in delays in the procedure, appear to be related to user error.It is likely that the guide wire was not pulled back before the stent deployment causing the stent to get trapped in the stent.Manipulation of the guide wire against resistance during retraction likely caused damage to the core resulting in the core separations, stretched coils and subsequent noted damages to the guide wire.The core was bent at the separation as if kinked prior to separation.There is no indication of a product quality issue with respect to manufacture, design, or labeling of the device.
|