| Model Number |
7308 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's age and weight were not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that she obtained blood glucose readings of 381 mg/dl at 7:09 a.M.And 99 mg/dl at 7:10 a.M.With the contour next ez meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next ez meter and contour next test strips from lot # 1cpeg09a for evaluation.Therefore, the returned meter was tested with the returned test strips using a blood sample, which gave a satisfactory performance.
|
| |
|