Model Number IPN045607
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Note: based on the additional information a second reported event is coded as serious injury.We had an incident at one of our facilities where the needles broke from two of these kits and one patient had to be taken to surgery to get it removed.We are in the process of getting the lot information and how much inventory is affected in the iu system but we will need to get replacement product sent asap.Additional information: the patient was discharged home in stable condition.She had a surgical removal of the catheter piece by our neurosurgeon.It was completely removed and discarded.There has not been a second patient to my knowledge that had it break while still in the patient.We have a couple that were difficult to remove but were removed fully intact and then broke externally in the crnas hands.
 
Event or Problem Description
Note: based on the additional information a second reported event is coded as serious injury.We had an incident at one of our facilities where the needles broke from two of these kits and one patient had to be taken to surgery to get it removed.We are in the process of getting the lot information and how much inventory is affected in the iu system but we will need to get replacement product sent asap.Additional information: the patient was discharged home in stable condition.She had a surgical removal of the catheter piece by our neurosurgeon.It was completely removed and discarded.There has not been a second patient to my knowledge that had it break while still in the patient.We have a couple that were difficult to remove but were removed fully intact and then broke externally in the crnas hands.
 
Additional Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter and epidural needle with no relevant findings.The ifu for this kit, e-17019-109a; rev.07, was reviewed as a part of this complaint investigation.The ifu warns the end user, "never advance catheter more than 5 cm beyond the needle tip.Advancing catheter more than 5 cm increases the likelihood of catheter-related complications." the ifu warns the user, "never tug or quickly pull on catheter during removal from patient to reduce risk of catheter breakage.Do not apply additional tension on the catheter if catheter begins to stretch excessively.Reposition patient to open the vertebral interspaces and re-attempt removal if resistance is encountered or if catheter stretches excessively during removal." a corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the epidural catheter and epidural needle with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.No further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key12647689
Report Number1036844-2021-00169
Device Sequence Number19156538
Product Code BSO
UDI-Device Identifier60801902005139
UDI-Public60801902005139
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberIPN045607
Device Catalogue NumberAK-05502
Device Lot Number23F21C0172
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2021
Supplement Date Received by Manufacturer11/21/2021
Initial Report FDA Received Date10/18/2021
Supplement Report FDA Received Date11/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-