| Model Number |
10618 |
| Medical Device Problem Codes |
Deflation Problem (1149); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It wa reported that difficulty removal and deflation failure occurred.The target lesion was located in the left anterior descending artery.A 4.50 x 16mm synergy ii drug eluting stent was advanced for treatment.After the stent was deployed in the lesion, the doctor tried to remove the stent balloon but it could not come out of the guiding catheter.The balloon was deflated for about 30 seconds a vacuum was held during removal after the stent would not deflate and was removed partially inflated.The physician remove the guiding catheter, wire and balloon as one unit to get everything out safely and completed the procedure with the same device.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It wa reported that difficulty removal and deflation failure occurred.The target lesion was located in the left anterior descending artery.A 4.50 x 16mm synergy ii drug eluting stent was advanced for treatment.After the stent was deployed in the lesion, the doctor tried to remove the stent balloon but could not come out of the guiding catheter after the stent would not deflate.So the physician remove the guiding catheter wire and balloon as one unit to get everything out safely and completed the procedure with the same device.There were no patient complications reported.It was further reported that the target lesion was 100% stenosed and was mildly tortuous and mildly calcified.
|
| |
|