• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3662
Device Problem Wireless Communication Problem (3283)
Patient Problems Failure of Implant (1924); Inadequate Pain Relief (2388)
Event Date 07/28/2021
Event Type  Injury  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event Description
It was reported that the patient's ipg stopped communicating with external devices following an unrelated surgery.A manufacturer representative met with the patient and deemed the ipg was inoperable.As a result, surgical intervention may take place at a later date to address the issue.
 
Manufacturer Narrative
A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
 
Event Description
Additional information received stated that the patient underwent surgical intervention wherein the ipg was explanted and replaced.Post-operatively, stimulation therapy was restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12651159
MDR Text Key277102579
Report Number1627487-2021-17769
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020222
UDI-Public05415067020222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model Number3662
Device Catalogue Number3662
Device Lot Number6179442
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received10/18/2021
Supplement Dates Manufacturer Received12/02/2021
01/18/2022
Supplement Dates FDA Received12/07/2021
02/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
Patient Weight91 KG
-
-