Model Number 3662 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problems
Failure of Implant (1924); Inadequate Pain Relief (2388)
|
Event Date 07/28/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Event Description
|
It was reported that the patient's ipg stopped communicating with external devices following an unrelated surgery.A manufacturer representative met with the patient and deemed the ipg was inoperable.As a result, surgical intervention may take place at a later date to address the issue.
|
|
Manufacturer Narrative
|
A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
|
|
Event Description
|
Additional information received stated that the patient underwent surgical intervention wherein the ipg was explanted and replaced.Post-operatively, stimulation therapy was restored.
|
|