| Model Number |
1012624-59 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The lot history record (lhr) for this product could not be reviewed and a similar complaint query could not be performed because the product was not returned for evaluation and the lot number was not reported.Based on the information provided, the reported difficulties and subsequent unexpected medical intervention appears to be due to circumstances of the procedure.The resistance during removal and stent dislodgement were likely due to interaction between the stent and introducer sheath during removal causing the stent to dislodge.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a left subclavian artery using a radial access point.The 8.0x59mm omnilink elite stent delivery system (sds) was advanced with no issue; however, once at the lesion the physician realized a smaller size stent was needed.The sds was attempted to be removed but resistance was met with the unspecified 6fr sheath and the stent became dislodged on the unspecified guide wire.A new unspecified 12fr sheath was inserted through the right groin and the new omnilink stent was successfully implanted.A 3.0 unspecified balloon was advanced and inflated to hold the dislodged 8.0x59mm omnilink elite stent in place on the balloon and removed successfully.There was no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|