U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH UNK - NAIL HEAD ELEMENTS: PFN BLADE; ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH UNK - NAIL HEAD ELEMENTS: PFN BLADE; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2021
Event Type  malfunction  
Manufacturer Narrative
Patient's birth year: 1953.This report is for an unk - nail head elements: pfn blade /unknown lot.Part and lot number are unknown.Without the specific part number; the udi number and 510-k number is unknown.Complainant part is not expected to be returned for manufacturer review/ investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that in (b)(6) 2020 a patient underwent the surgery to treat the subtrochanteric femur fracture with a marrow nail osteosynthesis (synthes proximal femoral nail (pfn) alpha).The patient suffers from slow progressive arthrosis and requested the nail construct to be removed despite clinically low symptoms.Pre-operative information about a possible one-sided procedure in case of later necessity of hip replacement has been provided; nevertheless, the patient wanted the material to be removed to make a possible future tep treatment easier.Material removal was done on (b)(6) 2021.During the procedure it was easy to find the blade end in the thigh, insert the guide wire and completely expose the blade end of scar tissue.While doing the attachment of the extraction screw via the guide wire (anti-clockwise); the thread broke off in the blade end without significant force, therefore the extraction screwdriver could no longer be fastened.The material removal had then to be canceled.The nail and blade remained in the patient.No fragments were created.The surgeon suspects a cold weld of nail and blade.Patient outcome is reported as good.Concomitant device reported: combined hammer 500g f/03.010.170 (part# 03.010.170, lot# 8843406, qty 1).Unk - cable/wire (part# unknown, lot# unknown, qty1).This complaint involves three (3) devices.This report is for one (1) unk - nail head elements: pfn blade.This is report 2 of 3 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - NAIL HEAD ELEMENTS: PFN BLADE
Type of Device
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12656568
MDR Text Key277147524
Report Number8030965-2021-08772
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
COMBINED HAMMER 500G F/03.010.170; EXTRACTSCR F/PFNA BLADE; UNK - CABLE/WIRE; UNK - NAILS: PFNA
Patient Outcome(s) Required Intervention;
-
-