• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23AGN-751
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/ method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
It was reported that a 23mm sjm regent heart valve was selected for an implant on (b)(6) 2021.One of the leaflet dislodged (as one piece) into the left ventricle (lv) after suturing and testing with the rubber tester.There was no resistance felt when the physician rotated the valve with the valve rotator.Device was explanted and when the physician removed the suturing, another leaflet also dislodged/ fall to the ground.There was a significant clinical delayed which could have been dangerous to patients life.A non abbott valve was implanted and the patient remained hemodynamically stable through the second implant.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information sections: d9, g3, h1, h2, h3, h6, h10.An event of the leaflet falling out when tested with the leaflet tester and the other falling out when the valve was removed was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key12659086
Report Number2648612-2021-00098
Device Sequence Number4852477
Product Code LWQ
UDI-Device Identifier05414734006101
UDI-Public05414734006101
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/02/2025
Device Model Number23AGN-751
Device Catalogue Number23AGN-751
Device Lot Number7761611
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2021
Supplement Date Received by Manufacturer11/11/2021
Initial Report FDA Received Date10/19/2021
Supplement Report FDA Received Date11/15/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-